1100 Series Nano Pump Service Manual 147
Repairing the Pump 3
Exchanging the Power Supply
1 Remove the pump from the stack, remove the front cover, top cover and top
foam section (see “Removing the Top Cover and Foam” on page 131).
2 Remove the EMPV. For this disconnect first:
the capillary going to the flow sensor.
the capillary coming from the filter.
the waste tube.
the connector.
3 Remove the flow sensor. For this disconnect first:
the capillary coming from the EMPV.
the capillary on the injection device (port 1).
the connector.
4 Unclip the mixing chamber from its holder.
5 Remove the damper and the Z-panel. For this disconnect first:
the mixing capillary.
the capillary going to the mixer.
the active inlet valve cable at the Z-panel.
the connector (P16, P11, P20) at the main board (see Figure 22 on page
135).
6 Remove both pump assemblies. For this, disconnect first:
the capillary coming from the mixing chamber.
the connecting tube coming from the solvent selection valve.
the connector (P8, P21 and P12, P14) at the main board.
Frequency If defective
Tools required Screwdriver Pozidriv #1
Wrench 1/4 inch
Wrench 14 mm
Wrench 7 mm
Wrench 5 mm
Parts required Power supply 0950-2528

Table of Contents

Other manuals for Agilent Technologies 1100 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1100 Series and is the answer not in the manual?

Agilent Technologies 1100 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model1100 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals