EasyManua.ls Logo

Agilent Technologies 1100 Series - Using the Detector

Agilent Technologies 1100 Series
288 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1100 Series Nano Pump Service Manual 45
Troubleshooting and Test Functions 2
Pump Configuration
At switch-on, the pump has recognized a new pump configuration.
The nano pump is assigned its configuration at the factory. If the active-inlet
valve and pump encoder of channel B are disconnected, and the nano pump is
rebooted, the error message is generated. However, the nano pump will
function as an isocratic pump in this configuration.The error message
reappears after each switch-on.
Probable Causes
Active-inlet valve and pump encoder of channel B disconnected.
Suggested Actions
Reconnect the active-inlet valve and pump encoder of channel B.

Table of Contents

Other manuals for Agilent Technologies 1100 Series

Related product manuals