100 1100 Series Nano Pump Service Manual
2 Troubleshooting and Test Functions
Wandering Detector Baseline
Critical Decision
Determine if the problem is in the DAD, or coming from the LC system:
Remove the flow cell from the DAD. Close the cell cover, and see if the baseline
performance improves.
1 If baseline performance has not improved:
a Replace the lamp(s).
b Evaluate the environment for excessive drafts, temperature changes, etc.
2 If baseline performance has improved, focus attention on the possible
causes and suggested actions below.
Table 12 Wandering Detector Baseline Possible Causes and Suggested Actions
Possible Causes Suggested Actions Notes
Dirty or defective flow cell Clean or replace the flow cell
The analytical column Bypass the column. Pump directly into the flow
cell. If performance improves, try a new column.
"Mixing Noise" when
pumping a binary mobile
phase from two channels.
Try premixing the mobile phase in one bottle,
and pumping 100% from that one solvent
channel.
If baseline performance improves, a solution
must be found which is a compromise between
mixing volume and other chromatographic
requirements. For an isocratic analysis,
premixing and pumping 100% one channel is the
best solution.
This problem occurs when one (or
both) of the solvents has a high
background absorbance at the
detection wavelength. In this case,
the pump’s mixing efficiency may
not be good enough to produce a
sufficiently homogeneous mobile
phase.
The detector reacts to gluts of the
more detectable parts of the solvent
mixture, and baseline disturbances
result.
Unstable flow and/or system
pressure.
Refer to the problem Unstable flow and/or
system pressure above. If your system suffers
from this problem, go through the possible
causes and suggested actions described there.
Failure to maintain stable column
flow or system pressure can also
cause unwanted baseline activity.

Table of Contents

Other manuals for Agilent Technologies 1100 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1100 Series and is the answer not in the manual?

Agilent Technologies 1100 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model1100 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals