1100 Series Nano Pump Service Manual 51
Troubleshooting and Test Functions 2
Encoder Missing
Encoder Missing: Pump channel A
B: Encoder Missing: Pump channel B
The optical encoder on the pump motor in the nano pump is missing or
defective.
The processor checks the presence of the pump encoder connector every 2
seconds. If the connector is not detected by the processor, the error message is
generated.
Probable Causes
Defective or disconnected pump encoder connector.
Defective pump drive assembly.
Suggested Actions
Ensure the connector is clean, and seated correctly.
Exchange the pump drive assembly.