1100 Series Nano Pump Service Manual 227
Introduction to the Nano Pump 5
Switches 1 and 2 do not force storage of this set of parameters in non-volatile
memory. Returning switches 1 and 2 to other positions (other than being both
up) will allow for normal operation.
If you use the following switch settings and power the instrument up again,
the instrument firmware stays in the resident part, that is, it is not operable as
a detector. It only uses basic functions of the operating system for example, for
communication.
To return to normal operation, set switches back to your GPIB or RS-232C
configuration settings.
Table 61 Stay Resident Settings
Mode Select12345678
TEST/BOOT11001000