EasyManua.ls Logo

Agilent Technologies 1100 Series - Status Indicators Explained

Agilent Technologies 1100 Series
288 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1100 Series Nano Pump Service Manual 97
Troubleshooting and Test Functions 2
Dirty EMPV (micro mode
only)
Perform the EMPV cleaning procedure, followed
by the EMPV performance test. Refer to the
Nano Pump Service Manual "EMPV Cleaning"
on page 91.
One of a capillary after the
EMPV is partially or
completely blocked
Remove the blocked capillary, clean it with
Acetone
Change the capillary
Any system component
which is offering a changing
restriction to the pump.
Replace the analytical column.
Replace the filter frit in front (upstream) of the
EMPV.
The vacuum micro-degasser
is off, or has become
defective.
Try another vacuum micro-degasser, or
experiment to determine the performance using
different degasser channels.
If the mobile phase is very sensitive to
gaseousness, use the micro-degasser
continuous mode.
Basic performance problems
in the pump.
Perform the pump Leak Test. Refer to the Nano Pump Service
Manual "Leak Test" on page 73.
Table 9 Unstable column flow and/or system pressure: Possible Causes and Suggested Actions, continued
Possible Causes Suggested Actions Notes

Table of Contents

Other manuals for Agilent Technologies 1100 Series

Related product manuals