74 1100 Series Nano Pump Service Manual
2 Troubleshooting and Test Functions
Ramp 2 The nano pump continues pumping solvent with a flow rate of 150 µl/min.
Channel A delivers for one pump cycle (first, plunger A2 delivers, then plunger
A1), followed by channel B (plunger B2, then plunger B1), both channels with
a stroke of 20 µl. The pressure increase during this phase should be linear.
Large leaks or defects will be characterized by an unstable, non-linear slope.
Ramp 3 Just before the start of the first plateau, plunger A2 delivers with a flow rate
of 50 µl/min for approximately 8 seconds. The system pressure should be 240
bar or higher.
Plateau 1 At plateau 1 plunger A2 delivers with a flow rate of 3 µl/min for 30 seconds.
During this time, the slope should be horizontal or slightly positive (slight
pressure increase). A negative slope indicates a leak rate greater than
3µl/min.
Ramp 4 Plunger B2 delivers 50 µl/min for approximately 8 seconds.
Plateau 2 Plunger B2 delivers with a flow rate of 3 µl/min for 30 seconds. The slope
should be horizontal or slightly positive. A negative slope indicates a leak rate
greater than 3 µl/min.
Ramp 5 Plunger A1 delivers 50 µl/min for approximately 8 seconds.
Plateau 3 Plunger A1 with a flow rate of 3 µl/min for 30 seconds. The slope should be
horizontal or slightly positive. A negative slope indicates a leak rate greater
than 3 µl/min.
Ramp 6 Plunger B1 delivers 50 µl/min for approximately 7 seconds.
Plateau 4 Plunger B1 delivers with a flow rate of 3 µl/min for approximately 30 seconds.
The slope should be horizontal or slightly positive. A negative slope indicates a
leak rate greater than 3 µl/min. At the end of the fourth plateau, the test is
finished and the nano pump switches off.
Running the Leak Test
Tools required Wrench 1/4 inch
Parts and materials
required
Restriction Capillary,G1313-87305
Blank nut, 01080-83202
Isopropanol, 500ml

Table of Contents

Other manuals for Agilent Technologies 1100 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1100 Series and is the answer not in the manual?

Agilent Technologies 1100 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model1100 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals