1220 Infinity II LC System User Manual 285
Maintenance and Repair
11
Autosampler
3
Remove the two fixing bolts from the base of the
metering head assembly.
4
Disassemble the metering head assembly.
5
Use a small screwdriver to carefully remove the seal.
Clean the chamber with lint-free cloth. Ensure all
particular matter is removed.
6
Install the new seal. Press the seal firmly into position.