Detectors
6850
6850
Released: April 2004 6850 Series Gas Chromatograph Service Procedures 193
4. Remove the exhaust tubing. 5. Remove the sheet metal cover. On the single
wavelength detector, it is held by two
screws at the top and two at the bottom,
while on the dual wavelength detector it is
held by two screws at the top.
Figure 1-26. Replacing the ignitor glow plug
Glow plug
Ignitor cable
O-ring
Spacer
Connector
Ignitor cable