Detectors
6850
6850
Released: April 2004 6850 Series Gas Chromatograph Service Procedures 194
6. Remove the ignitor cable.
7. Unscrew the connector and ignitor assembly
from the detector module.
8. Remove the O-ring, ignitor spacer, and glow
plug from the ignitor assembly. Replace the
glow plug. Inspect the O-ring and spacer, and
replace if needed.
9. Reassemble in reverse order.