3.3.1.4. Configured UE Transmitted Output Power
3.3.1.4.1. MT8820C
[Acceptable Value Setting]
1. [PCC/SCC] Execute PWR_AVG 20 to set the average count of Power measurement to 20 times.
2. [PCC/SCC] Execute TP_CONFPWR1_TOL 7.7 to set TX2 – Configured UE transmitted Output Power (Test
Point 1) Pass/Fail Judgment.
3. [PCC/SCC] Execute TP_CONFPWR2_TOL 6.7 to set TX2 – Configured UE transmitted Output Power (Test
Point 2) Pass/Fail Judgment.
4. [PCC/SCC] Execute TP_CONFPWR3_TOL 5.7 to set TX2 – Configured UE transmitted Output Power (Test
Point 3) Pass/Fail Judgment.
[Measurements]
5. [PCC/SCC] Execute TESTPRM TX_CONF_PWR1 to set Test Parameter to TX2 – Configured Power (Test
Point 1).
6. [PCC/SCC] Execute SWP to measure the power.
7. [PCC/SCC] Execute POWER? AVG to read the TX Power measurement result.
8. [PCC/SCC] Execute TESTPRM TX_CONF_PWR2 to set Test Parameter to TX2 – Configured Power (Test
Point 2).
9. Execute seps 6 to 7.
10. [PCC/SCC] Execute TESTPRM TX_CONF_PWR3 to set Test Parameter to TX2 – Configured Power (Test
Point 3).
11. Execute steps 6 to 7.
3.3.1.4.2. MT8821C
[Acceptable Value Setting]
1. Execute PWR_AVG 20 to set the average count of Power measurement to 20 times.
2. Execute TP_CONFPWR1_TOL 7.7 to set TX2 – Configured UE transmitted Output Power (Test Point 1)
Pass/Fail Judgment.
3. Execute TP_CONFPWR2_TOL 6.7 to set TX2 – Configured UE transmitted Output Power (Test Point 2)
Pass/Fail Judgment.
4. Execute TP_CONFPWR3_TOL 5.7 to set TX2 – Configured UE transmitted Output Power (Test Point 3)
Pass/Fail Judgment.
[Measurements]
5. Execute TESTPRM TX_CONF_PWR1 to set Test Parameter to TX2 – Configured Power(Test Point 1).
6. Execute SWP to measure the power.
7. Execute POWER? AVG,PCC to read the PCC TX Power measurement result.
8. Execute POWER? AVG,SCC1 to read the SCC–1 TX Power measurement result.
9. Execute TESTPRM TX_CONF_PWR2 to set Test Parameter to TX2 – Configured Power(Test Point 2).
10. Execute steps 6 to 8.
11. Execute TESTPRM TX_CONF_PWR3 to set Test Parameter to TX2 – Configured Power(Test Point 3).
12. Execute steps 6 to 8.
NOTE 1: The tolerance (lower limit) is relaxed by 1.5dB for transmission bandwidths confined within F
UL_low
and F
UL_low
+ 4 MHz or F
UL_high
– 4 MHz and F
UL_high
.
NOTE 2: The relief requirement ΔT
IB,c
is applied to each band for Inter–band CA
Refer to TS36.521–1 Table 6.2.5.3-2 for ΔT
IB,c
.
The following conditions are applied to compatible UEs with more than one Inter–band CA
Configuration.
For carrier frequency f ≤ 1 GHz: mean value of ΔT
IB,c
in compatible Band Configuration
For carrier frequency > 1 GHz: maximum value of ΔT
IB,c
in compatible Band Configuration