119
3.3.1.5. Minimum Output Power
3.3.1.5.1. MT8820C
[Acceptable Value Setting]
1. [PCC/SCC] Execute PWR_AVG 20 to set the average count of Power measurement to 20 times.
2. [PCC/SCC] Execute TP_MINPWR_UL -39.0 to set TX1 – Min. Power Pass/Fail judgment.
[Measurements]
3. [PCC/SCC] Execute TESTPRM TX_MINPWR to set Test Parameter to TX1 – Min. Power.
4. [PCC/SCC] Execute SWP to measure the power.
5. [PCC/SCC] Execute CHPWR? AVG to read the Channel Power measurement result.
6. [PCC/SCC] Execute CHPWRPASS? to check that the Channel Power measurement Pass/Fail judgment is Pass.
3.3.1.5.2. MT8821C
[Acceptable Value Setting]
1. Execute PWR_AVG 20 to set the average count of Power measurement to 20 times.
2. Execute TP_MINPWR_UL -39.0 to set TX1 – Min. Power Pass/Fail judgment.
[Measurements]
3. Execute TESTPRM TX_MINPWR to set Test Parameter to TX1 – Min. Power.
4. Execute SWP to measure the power.
5. Execute CHPWR? AVG,PCC to read the PCC Channel Power measurement result.
6. Execute CHPWRPASS? PCC to check that the PCC Channel Power measurement Pass/Fail judgment is Pass.
7. Execute CHPWR? AVG,SCC1 to read SCC–1 Channel Power measurement result.
8. Execute CHPWRPASS? SCC1 to check that the SCC–1 Channel Power measurement Pass/Fail judgment is
Pass.
NOTE 1: Pass/Fail judgment values differ depending on the Carrier Frequency f.
f ≤ 3.0GHz : ≤ –39 dBm (initial value)
3.0GHz < f ≤ 4.2GHz : ≤ –38.7 dBm
3.3.1.6. UE Transmit OFF Power
Refer to Chapter 3.3.1.7