120
3.3.1.7. General ON/OFF Time Mask
3.3.1.7.1. MT8820C
[Acceptable Value Setting]
1. [PCC/SCC] Execute TP_OFFPWR_UL -48.5 to set TX2 – General Time Mask Off Power Pass/Fail judgment.
2. [PCC/SCC] Execute TP_TMASK_GEN_TOL 7.5 to set TX2 – General Time Mask On Power Pass/Fail
judgment.
[Measurements]
3. [PCC/SCC] Execute TESTPRM TX_GEN_TMASK to set Test Parameter to TX2 – General Time Mask.
4. [PCC/SCC] Execute PT_WDR ON to enable Power Template Wide Dynamic Range measurement.
5. [PCC/SCC] Execute SWP to measure the Power Template.
6. [PCC/SCC] Execute ONPWR? AVG to read the On Power measurement result.
7. [PCC/SCC] Execute ONPWRPASS? to check that the On Power measurement Pass/Fail judgment is Pass.
8. [PCC/SCC] Execute OFFPWR_BEFORE? AVG to read the Off Power (Before) measurement result.
9. [PCC/SCC] Execute OFFPWR_AFTER? AVG to read the Off Power (After) measurement result.
10. [PCC/SCC] Execute OFFPWRPASS? to check that the Off Power measurement Pass/Fail judgment is Pass.
3.3.1.7.2. MT8821C
[Acceptable Value Setting]
1. Execute TP_OFFPWR_UL -48.5 to set TX2 – General Time Mask Off Power Pass/Fail judgment.
2. Execute TP_TMASK_GEN_TOL 7.5 to set TX2 – General Time Mask On Power Pass/Fail judgment.
[Measurements]
3. Execute TESTPRM TX_GEN_TMASK to set Test Parameter to TX2 – General Time Mask.
4. Execute PT_WDR ON to enable Power Template Wide Dynamic Range measurement.
5. Execute SWP to measure the Power Template.
6. Execute ONPWR? AVG,PCC to read the PCC On Power measurement result.
7. Execute ONPWRPASS? PCC to check that the PCC On Power measurement Pass/Fail judgment is Pass.
8. Execute OFFPWR_BEFORE? AVG,PCC to read the PCC Off Power (Before) measurement result.
9. Execute OFFPWR_AFTER? AVG,PCC to read the PCC Off Power (After) measurement result.
10. Execute OFFPWRPASS? PCC to check that the PCC Off Power measurement Pass/Fail judgment is Pass.
11. Execute ONPWR? AVG,SCC1 to read the SCC–1 On Power measurement result.
12. Execute ONPWRPASS? SCC1 to check that the the SCC–1 On Power measurement Pass/Fail judgment is Pass.
13. Execute OFFPWR_BEFORE? AVG,SCC1 to read the SCC–1 Off Power (Before) measurement result.
14. Execute OFFPWR_AFTER? AVG,SCC1 to read the SCC–1 Off Power (After) measurement result.
15. Execute OFFPWRPASS? SCC to check that the the SCC–1 Off Power measurement Pass/Fail judgment is Pass.
NOTE 1: Pass/Fail judgment values for Transmitted Off Power differ depending on the Carrier Frequency f
f ≤ 3.0GHz : ≤ –48.5 dBm (initial value)
3.0GHz < f ≤ 4.2GHz : ≤ –48.2 dBm