3.3.1.8. Power Control Absolute Power Tolerance
3.3.1.8.1. MT8820C
[Acceptable Value Setting]
1. [PCC/SCC] Execute TP_PCTABS_TOL 10.0 to set TX3 – Absolute Power (Test Point1) Pass/Fail judgment.
[Measurements]
2. [PCC/SCC] Execute TESTPRM TX_PCTABS1 to set Test Parameter toTX3 – Absolute Power (Test Point1).
3. [PCC/SCC] Execute SWP to measure the Power Control Tolerance (Absolute Power).
4. [PCC/SCC] Execute PCTPWR? to read the Absolute Power (dBm) measurement result.
5. [PCC/SCC] Execute PCTPASS? to check that the Absolute Power measurement Pass/Fail judgment is Pass.
6. [PCC/SCC] Execute TESTPRM TX_PCTABS2 to set Test Parameter to TX3 – Absolute Power (Test Point2).
7. Execute step 3 to 5.
3.3.1.8.2. MT8821C
[Acceptable Value Setting]
1. Execute TP_PCTABS_TOL 10.0 to set TX3 – Absolute Power (Test Point1) Pass/Fail judgment.
[Measurements]
2. Execute TESTPRM TX_PCTABS1 to set Test Parameter to TX3 – Absolute Power (Test Point1).
3. Execute SWP to measure the Power Control Tolerance (Absolute Power).
4. Execute PCTPWR? PCC to read the PCC Absolute Power (dBm) measurement result.
5. Execute PCTPASS? PCC to check that the PCC Absolute Power measurement Pass/Fail judgment is Pass.
6. Execute PCTPWR? SCC1 to read the SCC–1 Absolute Power (dBm) measurement result.
7. Execute PCTPASS? SCC1 to check that the SCC–1 Absolute Power measurement Pass/Fail judgment is Pass.
8. Execute TESTPRM TX_PCTABS2 to set Test Parameter to TX3 – Absolute Power (Test Point2).
9. Execute step 3 to 7.
NOTE 1: The Pass/Fail judgment value for Expected Measured Power differs depending the Carrier
Frequency f.
f ≤ 3.0GHz : ≤ 10.0 dBm (as the initial value)
3.0GHz < f ≤ 4.2GHz : ≤ 10.4 dBm