3.3.2.14. In-band emissions for non allocated RB for CA (intra-band contiguous DL CA and UL CA)
(6.5.2A.3-1)
3.3.2.14.1. MT8820C
This measurement item is not supported by the MT8820C.
Refer to Table 3.31.
3.3.2.14.2. MT8821C
This subsection describes an example of intraband measurement.
Example: PCC N
RB
= 100, SCC N
RB
= 50, N
RB_alloc
= 12, Modulation = QPSK
PCC and SCC RB allocations (L
CRB
@RB
start
) are P_12@0 and S_0@0, respectively
For Intraband measurement, the Carrier Leakage Frequency Parameter must be set properly set before the start
of measurement sequence. For more information, see Annex B.2.
[Carrier Leakage Frequency Setting]
1. Execute IBEM_CLFR CFR to set Carrier Leakage Frequency to at Carrier Frequency Center.
[Pass/Fail evaluation limits value setting]
2. Execute MOD_AVG 20 to set the average count of Modulation Analysis to 20 times.
3. Execute TP_INBANDE_GEN_A -29.2 (or TP_INBANDE_GEN_A -29.2, PCC) to set General Pass/Fail judgment of
TX1 - IBE/LEAK @ 0/30/40 dBm for PCC.
4. Execute TP_INBANDE_GEN_B -24.2 (or TP_INBANDE_GEN_B -24.2, PCC) to set General Pass/Fail judgment of
TX1 - IBE/LEAK @ 0/30/40 dBm for PCC.
5. Execute TP_INBANDE_GEN_C -2.2 (or TP_INBANDE_GEN_C -2.2, PCC) to set General Pass/Fail judgment of
TX1 - IBE/LEAK @ 0/30/40 dBm for PCC.
6. Execute TP_INBANDE_GEN_D -56.2 (or TP_INBANDE_GEN_D -56.2, PCC) to set General Pass/Fail judgment of
TX1 - IBE/LEAK @ 0/30/40 dBm for PCC.
7. Execute TP_INBANDE_IMG 24.2 (or TP_INBANDE_IMG 24.2, PCC) to set General Pass/Fail judgment of TX1
- IBE/LEAK @ 0/30/40 dBm for PCC.
8. Execute TP_INBANDE_LEAK_0DBM 24.2 (or TP_INBANDE_LEAK_0DBM 24.2, PCC) to set General Pass/Fail
judgment of TX1 IBE/LEAK @ 0 dBm for PCC.
9. Execute TP_INBANDE_LEAK_M30DBM 19.2 (or TP_INBANDE_LEAK_M30DBM 19.2, PCC) to set General
Pass/Fail judgment of TX1 IBE/LEAK @ 30 dBm for PCC.
10. Execute TP_INBANDE_LEAK_M40DBM 9.2 (or TP_INBANDE_LEAK_M40DBM 9.2, PCC) to set General
Pass/Fail judgment of TX1 IBE/LEAK @ 40 dBm for PCC.
11. Execute TP_INBANDE_GEN_A -29.2, SCC1 to set General Pass/Fail judgment of TX1 IBE/LEAK @ 0/30/40
dBm for SCC1.
12. Execute TP_INBANDE_GEN_B -24.2, SCC1 to set General Pass/Fail judgment of TX1 IBE/LEAK @ 0/30/40
dBm for SCC1.
13. Execute TP_INBANDE_GEN_C -2.2, SCC1 to set General Pass/Fail judgment of TX1 IBE/LEAK @ 0/30/40
dBm for SCC1.
14. Execute TP_INBANDE_GEN_D -57.0, SCC1 to set General Pass/Fail judgment of TX1 IBE/LEAK @ 0/30/40
dBm for SCC1.
15. Execute TP_INBANDE_IMG 24.2, SCC1 to set General Pass/Fail judgment of TX1 IBE/LEAK @ 0/30/40
dBm for SCC1.
16. Execute TP_INBANDE_LEAK_0DBM 24.2, SCC1 to set General Pass/Fail judgment of TX1 IBE/LEAK @ 0
dBm for SCC1.
17. Execute TP_INBANDE_LEAK_M30DBM 19.2, SCC1 to set General Pass/Fail judgment of TX1 IBE/LEAK @
30 dBm for SCC1.
18. Execute TP_INBANDE_LEAK_M40DBM 9.2, SCC1 to set General Pass/Fail judgment of TX1 IBE/LEAK @ 40
dBm for SCC1.
[Measurements for test configuration of first example]
19. Execute TESTPRM TX_0DBM to set Test Parameter to TX1 IBE/LEAK @ 0 dBm.
20. Execute ULRMC_RB 12 to set Common Parameter UL RMC Number of RB to 12.
(For other Configuration ID setting, set Common Parameter UL RMC Number of RB and Starting RB
according to TS36.5211 Table 6.5.2A.3.1.4.11.)
21. Execute ULRMC_RB_SCC1 0 to set Common Parameter SCC-1 UL RMC - Number of RB to 0.
22. Execute SWP to measure the Modulation Analysis.
23. Execute INBANDE_GEN? MAX (or INBANDE_GEN? MAX, PCC) to read the Inband Emissions (General)
measurement result of PCC (allocated component carrier).
24. Execute INBANDE_IMG? MAX (or INBANDE_IMG? MAX, PCC) to read the Inband Emissions (IQ Image)

Table of Contents