2.2.3.2. MT8821C
2.2.3.2.1. Setting Example 1 (Intra-Band FDD DL CA and UL CA)
This chapter describes a setting example for Duplex Mode set to FDD, Intra-Band DL CA and UL CA.
Set both Test Channel Bandwidth PCC and SCC to 20 MHz.
6. Execute PRESET to initialize parameters.
7. Execute CHCODING RMC_DLUL_CA_PCC to set Common Parameter - Channel Coding to RMC(DL/UL CA).
8. Execute CALLPROC ON to set Common Parameter - Call Processing to ON.
9. Execute DLCHAN 200 to set Common Parameter - Frequency - DL Channel to 200 simultaneously with UL
Channel to 18200.
10. Execute DLCHAN_SCC1 398 to set Common Parameter - SCC-1 - DL Channel to 398.
11. Execute BANDWIDTH 20MHZ to set Common Parameter - Channel Bandwidth to 20 MHz.
12. Execute BANDWIDTH_SCC1 20MHZ to set Common Parameter - SCC-1 - Channel Bandwidth to 20 MHz.
2.2.3.2.2. Setting Example 2 (Inter-Band FDD DL CA without UL CA)
This chapter describes a setting example for Duplex Mode set to FDD, Inter-Band DL CA without UL CA.
1. Execute PRESET to initialize parameters.
2. Execute CHCODING RMC_DL_CA_PCC to set Common Parameter - Channel Coding to RMC(DL CA).
3. Execute CALLPROC ON to set Common Parameter - Call Processing to ON.
4. Execute DLCHAN 200 to set Common Parameter - Frequency - DL Channel to 200 simultaneously with UL
Channel to 18200.
5. Execute DLCHAN_SCC1 2525 to set Common Parameter - SCC-1 - DL Channel to 2525.
6. Execute BANDWIDTH 10MHz to set Common Parameter - Channel Bandwidth to 10 MHz.
7. Execute BANDWIDTH_SCC1 10MHZ to set Common Parameter - SCC-1 - Channel Bandwidth to10 MHz.
2.2.3.2.3. Setting Example 3 (Intra-Band TDD DL CA without UL CA)
This chapter describes a setting example for Duplex Mode set to TDD, Intra-Band DL CA without UL CA.
The Uplink/Downlink Configuration is set to 1, and Special Subframe Configuration is set to 4.
Follow the procedure in Chapter 2.2.3.2.2 replacing Step 4, 5 and followed by Step 6 and 7 as below.
4. Execute DLCHAN 38000 to set Common Parameter - DLChannel simultaneously with UL Channel to 38000.
5. Execute DLCHAN_SCC1 38099 to set Common Parameter - SCC-1 - DL Channel to 38099.
6. Execute TDDULDLCONF 1 to set Common Parameter - TDD Uplink/Downlink Configuration to 1.
7. Execute TDDSSFCONF 4 to set Common Parameter - TDD Special Subframe Configuration to 4.
NOTE 1: Set same value as Uplink/Downlink Configuration in PCC and SCC.
NOTE 2: Set same value as Special Subframe Configuration in PCC and SCC.
2.2.3.2.4. Setting Example 4 (Inter Band TDD DL CA and UL CA)
This chapter describes a setting example for Duplex Mode set to TDD, Inter-Band DL CA and UL CA.
The Uplink/Downlink Configuration is set to 2, and Special Subframe Configuration is set to 5.
Follow the procedure in Chapter 2.2.3.2.1 replacing with Step 4, 5 and followed by Step 6 and 7 as below.
4. Execute DLCHAN 38000 to set Common Parameter - DL Channel simultaneously with UL Channel to
38000.
5. Execute DLCHAN_SCC1 39150 to set Common Parameter - SCC-1 - DL Channel to 39150.
6. Execute TDDULDLCONF 2 to set Common Parameter - TDD Uplink/Downlink Configuration to 2.
7. Execute TDDSSFCONF 5 to set Common Parameter - TDD Special Subframe Configuration to 5.
NOTE 1: Set same value as Uplink/Downlink Configuration in PCC and SCC.
NOTE 2: Set same value as Special Subframe Configuration in PCC and SCC.