2-46
AX-4030
AUTION MAX AX-4030 OPERATING MANUAL
When check measurement is completed, the
obtained result will be reported on the screen.
5 Press Print to print the result.
6 Press Go back to return to the [Maintenance] screen.
6 Measure the second test strip (gray strip).
1 Press 3 to go to the [Check measurement] screen.
2 Select the same feeder as the one for the first strip.
For detailed instructions, see step 3 .
3 Take a gray test strip out of the container.
4 Load the gray test strip into the feeder selected in
step 2.
For detailed instructions, see step 5 .
Check measurement
(5300) 2006-01-09 14:11
Measurement completed.
***** 2006-01-09 14:10 ****
430[nm] 89.3 %
500[nm] 93.1 %
565[nm] 88.5 %
635[nm] 9.7 %
760[nm] 100.0 %
Print Go back
ReflectivityWavelengths
Maintenance
(5000) 2006-01-09 14:15
1 S.G. cell washout
2 Clean washing bath and tray
3 Check measurement
4 Maintenance information
Check measurement
(5300) 2006-01-09 14:15
Feeder No.
[1]
OK Go back
Gray check strip
Loading slot
Black marker
Black marker facing down
Gray check strip

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Arkray AUTION MAX AX-4030 and is the answer not in the manual?

Arkray AUTION MAX AX-4030 Specifications

General IconGeneral
BrandArkray
ModelAUTION MAX AX-4030
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals