Programming
65
5. Tuning factor Used in “DS” tool for the torque tuning factor. Also used in
“PF3000” with tool model “Tensor DS”.
6. Communication Used in tool calibration to select the type of communication that is
to be used. For each type of controller, “Manual input” or
“RS232” can be selected. For DS controller and PF3000,
”Infrared” can also be selected. When selecting “Manual input”, the
user is asked for controller torque and controller angle in input
fields after each tightening operation. When “RS232” or “Infrared”
is selected, controller tightening values are automatically read from
the controller after each tightening operation.
7. Parameter set Used in “F2000/PF2000”, “PF3000” and “DS” tool calibration for
selecting the parameter set to be used when calibrating.
8. Cell ID/Unit ID Used in PF3000 tool calibration to select cell and in
Focus2000/PF2000 to select unit ID.
9. Channel Used in Focus2000/PF2000 and PF3000 tool calibration to select
channel.
10. Model Used in PF3000 tool calibration to select tool model. You can
choose between Tensor S or Tensor DS.