10.3.13.2 Metering monitoring / Stop pump
Fig. 10.35 Metering monitoring / Stop pump
The “stop pump” function is used
to set whether or not the pump
should be stopped if a “metering
monitoring alarm” is triggered.
- Pump is not stopped when a
“metering
monitoring alarm”
is triggered. (Default setting)
Pump is stopped
when a “metering
monitoring alarm”
is triggered.
10.3.13.3 Metering monitoring / Strokes
Fig. 10.36 Metering monitoring / Strokes
The “strokes” option indicates the
stroke interval to be monitored.
Setting range: 0 – 100
Default setting = 10
10.3.13.4 Metering monitoring / Deviation (only at E60
PLUS
& OGM or OGM
PLUS
)
A tolerance in % is set under “deviation”.
The actual incoming pulse rate recorded by the oval gear meter within the pump interval to be
monitored (see chapter
10.3.13.3 “Configuration/Strokes”) is compared with the target pulse
rate defined during calibration.
If this comparison results in a deviation that is greater than the % value configured under
“deviation”, a “metering monitoring alarm” is triggered.
NOTE
This function can only be selected if a dongle box is connected and configuration/oval
gear meter (see chapter 10.3.11) has been activated.
Fig. 10.37 Metering monitoring / Deviation (only at E60
PLUS
&
OGM or OGM
PLUS
Setting range: 0-100%
Default setting = 20 %