Fit the dongle- box or the MicroFlow- Box (Fig. 8.1) to the rails of the EMP pump provided for
this purpose (Fig. 8.2, Pos. A):
Slide the dongle box (Fig. 8.1) onto the rail on the left side of the EMP pump
(Fig. 8.2, Pos. A), until it snaps into place.
Connect the connection cable (Fig. 8.1, Pos. 3 and 8.3, Pos. 2) to the input for pulse /
standard signal (Fig. 8.2, Pos. 1).
Switch the pump off and afresh on.
The additional functions in the metering pump electronics can thus be activated and
executed.
NOTE
If, before upgrading with a dongle box or a MicroFlow-box, a plug is connected to the input
for pulse / standard signal (see Fig. 8.2, Pos. 1), then this plug can be connected after the
upgrade to plug VIII (see Fig. 8.1, Pos. VIII) of the dongle box or Microflow-box.
8.2 Connector assignments of the slots
8.2.1 Connector assignments of slot V (5-terminal), Dongle-box
Pin
Conductor coloring
(cable connection)
Assignment
Supply for oval gear meter 5 V DC (load with max. 50 mA)
Bus line for oval gear meter, type OGM
PLUS
Fig. 8.4 Connector assignments of slot V (5-terminal), input for Dongle-box
Remove the protective cap
Mount the connector plug according to the connector assignment.
WARNING
Only use a connector plug from our product range!
8.2.1.1 Installing the oval gear meter OGM
PLUS
1 + 2 + 3+ 4 = OGM
Fig. 8.5 Installing the oval gear meter OGM
PLUS
8.2.2 Connector assignments of Slot V, MicroFlow- Box
Pin
Conductor coloring
(cable connection)
Assignment
Signal input MicroFlow- Transducer
Fig. 8.6 Connector assignments of slot V (5-terminal), MicroFlow-Box