12.2.1 Deactivating of Dongle Box, OGM
PLUS
, MicroFlow
PLUS
in the pump software
The following procedure is applied to re-disable a peripheral unit (dongle box, OGM
PLUS
,
MicroFlow
PLUS
), which had previously been connected and has then been removed again, in
the pump software.
Cancel the alarm signal which appears by pressing the Start/Stop button.
The following appears in the display unit:
Fig. 12.15 Operating faults:
Deactivating of DongleBox in the pump
Fig. 12.16 Operating faults:
Deactivating of OGM
PLUS
in the pump
Fig. 12.17 Operating faults:
Deactivating of MicroFlow
PLUS
in the pump
Deactivation can be selected by pressing the or button and
the following is displayed
Fig. 12.18 Operating faults:
Deactivating of DongleBox in the pump
Fig. 12.19 Operating faults:
Deactivating of OGM
PLUS
in the pump
Fig. 12.20 Operating faults:
Deactivating of MicroFlow
PLUS
in the pump
The setting is taken over through pressing the Start/Stop button
12.3 Trouble shooting
metering pump does not
work, no display indicator
mains power cable damaged
despite venting and max.
deposits, adhesions or drying-out
of the valves
rinse the metering head through the
suction line, if necessary remove and
clean or replace the valves
medium escapes from the
diaphragm breakage outlet
metering head is loose
screw in the metering head fastening
screw diagonally
low-level indicator
appears on the display
despite a full container
float of the suction pipe is jammed
suction pipe plug or strapping plug
is loose or not plugged in
tighten the plug, clean the contacts,
check whether the strapping plug is
suction pipe cable is faulty
replace the empty report device
CAUTION
Prior to repair and maintenance work and metering of dangerous media, always rinse the
metering head, relieve the pressure pipe and wear protective clothing (protective goggles,
Electronics repairs must only be performed by trained electricians, following the safety
regulations of the professional association VB G 4 and ZH 1/11)!
CAUTION
When opening the covers or removing parts, except when this is possible without tools,
voltage-carrying parts may be exposed. Connection points may also be under live
Before calibration, maintenance, repairs or replacement of parts, the device must be
disconnected from all voltage sources if it is necessary to open up the device.