10.3.14.1 Selecting
Fig. 10.40 Configuration MicroFlow
PLUS
: Selecting
- MicroFlow
PLUS
is inactive.
(Default setting)
MicroFlow
PLUS
is active.
On activating MicroFlow
PLUS
, a
submenu appears with the following
menu items:
Menu item see chapter
Stop pump 10.3.14.2
Strokes
10.3.14.3
Deviation
10.3.14.4
Metering break
10.3.14.5
Metering delay
10.3.14.6
Calibration 10.3.14.7
10.3.14.2 MicroFlow
PLUS
/ Stop pump
Fig. 10.41 Configuration MicroFlow
PLUS
/ Stop pump
The “stop pump” function is used
to set whether or not the pump
stops if a “metering monitoring
alarm” is triggered.
- Pump is not stopped in the
event of anAlarm metering
monitor function
(Default setting)
Pump is stopped in the event
of anAlarm metering monitor
function
10.3.14.3 MicroFlow
PLUS
/ Strokes
Fig. 10.42 Configuration MicroFlow
PLUS
/ Strokes
The "Strokes" option is used to
specify the interval of strokes to
be monitored.
Setting range: 0-100
Default setting = 10 strokes