7.2.3.5 Installing the metering monitoring
NOTE
The metering monitoring function must be activated in menu first. The selection must also
be made of whether the pins 2+4 (input “pulse”) or 3+4 (input “metering lock”) should be
used for the metering monitoring input (see chapter 10.3.13).
Fig. 7.17 Installing the metering monitoring
2 + 4 = Input metering monitoring function with potential-free contact with selection input „pulse“
at configuration / metering monitoring (see chapter 10.3.13).
1 +2 + 4 = Input metering monitoring function with electronic switch (NPN) with selection input
„pulse“ at configuration / metering monitoring (see chapter 10.3.13).
or
3 + 4 = Input metering monitoring function with potential-free contact with selection input
„metering lock“ at configuration / metering monitoring (see chapter 10.3.13).
1 +3 + 4 = Input metering monitoring function with electronic switch (NPN) with selection input
„metering lock“ at configuration / metering monitoring (see chapter 10.3.13).
7.2.4 Connector assignments of slot III (4-terminal) output for alarm and stroke signal
Pin
Conductor coloring
(cable connection)
Assignment
Output low-level warning, empty report and fault (+)
Output low-level warning, empty report and fault (-)
(plug assignments: see chapter 14.3.5.3 “Pin assignment / conductor coloring Connector III
(4-terminal)“)
Fig. 7.18 Connector assignments of slot III (4-terminal) output for alarm and stroke signal
Remove the protective cap
Mount the connector plug according to the connector assignment.
WARNING
Only use a connector plug from our product range! (included in the delivery scope)