Patient Monitor User Manual Monitoring NIBP
- 121 -
pressure against the cuff.
2. Connect the air hose to the connector on XM module and switch on the monitor.
Apply the blood pressure cuff to the patient's arm or leg and follow the instructions below.
-Ensure that the cuff is completely deflated.
-Apply the appropriate size cuff to the patient (About the cuff size selection, please refer to
Section NIBP Accessories), and make sure that the symbol "Φ" is over the artery. Ensure that the
cuff is not wrapped too tightly around the limb. Excessive tightness may cause discoloration and
eventual ischemia of the extremity.
Cuff Usage
3. Check whether the patient type is appropriately selected. Access the Patient Setup menu
from Menu. Turn the knob to select the required patient Type in the Patient Info. menu.
4. Select a measurement mode in the NIBP Setup menu. Refer to section Operation Prompts
for details.
5. Press the button on the front panel or shortcut key on the screen to start a
measurement.
6. Wait until the first reading is taken.
NOTE:
1 The width of the cuff is either approximately 40% of the limb circumference or 2/3 of
the upper arm length. The inflatable part of the cuff should be long enough to encircle
80%-100% of the limb. The wrong size of cuff can cause erroneous readings. If the
cuff size is in question, use another cuff with suitable size to avoid errors.
2 If an NIBP measurement is suspect, repeat the measurement. If you are still
uncertain about the reading, use another method to measure the blood pressure.
3 Please make sure the cuff is well connected. A leak of air may cause measurement
error.
4 Please select the cuff with the suitable size. An unsuitable cuff may cause incorrect
measurements.
5 Avoid incursion of liquid into the cuff. If this happens, please desiccate the cuff
completely.