A.13 NIBP
Complies with IEC 80601-2-30: 2009+A1: 2013.
EDAN Module
Technique Oscillometry
Mode Manual, Auto, Continuous
Measuring Interval in AUTO
Mode (unit: minute)
1/2/3/4/5/10/15/30/60/90/120/180/240/360/480
Continuous 5 min, interval is 5 s
Measuring Type SYS, DIA, MAP, PR
☆Measuring Range
☆Adult Mode
SYS: 40 mmHg to 270 mmHg
DIA: 10 mmHg to 215 mmHg
MAP: 20 mmHg to 235 mmHg
☆Pediatric Mode
SYS: 40 mmHg to 230 mmHg
DIA: 10 mmHg to 180 mmHg
MAP: 20 mmHg to 195 mmHg
☆Neonatal Mode
SYS: 40 mmHg to 135 mmHg
DIA: 10 mmHg to 100 mmHg
MAP: 20 mmHg to 110 mmHg
☆Alarm Type
SYS, DIA, MAP
☆ Cuff Pressure Measuring
Range
0 mmHg to 300 mmHg
Pressure Resolution 1 mmHg
☆Maximum Mean Error
± 5 mmHg
☆Maximum Standard Deviation
8 mmHg
Maximum Measuring Period
Adult/Pediatric 120 s
Neonate 90 s
Typical Measuring Period 20 s to 35 s (depend on HR/motion disturbance)
Dual Independent Channel Overpressure Protection
Adult (297 ± 3) mmHg
Pediatric (245 ± 3) mmHg