Patient Monitor User Manual Alarms
- 44 -
4.3.2 Audio Alarm Paused
You can temporarily prevent alarms from sounding by pressing the hardkey on the front
panel or pressing shortcut key on the screen.
You can set the alarm pause time as desired. The default alarm pause time is 120 s.
1. Select Menu > Maintenance > User Maintain, and enter the required password ABC.
2. Select Alarm Setup, and set Pause Time to 60 s, 120 s, or 180 s.
When alarms are paused,
The audio alarm is turned off, and no alarms are sounding.
The visual alarm indications are still displayed.
The monitor displays the audio alarm paused icon .
The monitor displays the remaining pause time in seconds with red background.
The hardkey on the front panel flashes in yellow.
When the alarm pause time expires, the audio alarm paused status is automatically terminated and
alarm is sounding. You can also terminate the alarm paused status by pressing the hardkey
.
NOTE:
If a new alarm occurs during the audio alarm paused period, the new alarm will not be
sounding.
4.3.3 Audio Alarm off
Set Pause Time to Permanent, press hardkey or shortcut key , the monitor
displays information: please confirm whether to activate audio alarm off function? Click Yes,
the monitor will enter into audio alarm off status. Click No, the monitor will keep current status.
During the audio alarm off status,
The audio alarm is turned off, and no alarms are sounding.
The visual alarm indications are still displayed.
The hardkey
on the front panel flashes in yellow.
Remind signal: Audio alarm off symbol and Audio Alarm off on a red colored
background are displayed with an interval of 2 s during the audio alarm off status. If module
loading or data transferring is in progress at the meantime, the remind signal for audio alarm off
will disappear till the module loading or data transferring is finished.