Patient Monitor User Manual Default Settings
- 291 -
Alarm High Limit (DIA) 90 70 60
Alarm Low Limit (DIA) 50 40 20
Inflation value EDAN Module 160 140 100
Omron Module 180 180 120
SunTech Module 160 140 90
Unit mmHg
Interval Manual
C.8 TEMP Default Settings
TEMP Settings ADU PED NEO
Alarm Switch On
Alarm Record Off
Alarm Level Medium
Alarm High Limit (T1) 39.0 39.0 39.0
Alarm Low Limit (T1) 36.0 36.0 36.0
Alarm High Limit (T2) 39.0 39.0 39.0
Alarm Low Limit (T2) 36.0 36.0 36.0
Alarm High Limit (TD) 2.0 2.0 2.0
Unit
°C
C.9 IBP Default Settings
IBP Settings ADU PED NEO
Alarm Switch On
Alarm Record Off
Alarm Level Medium
Unit mmHg
Filter 12.5 Hz
SYS/DIA/MAP SYS/DIA/MAP SYS/DIA/MAP
Alarm High Limit (ART, P1, P2) 160, 90, 110 120, 70, 90 90, 60, 70
Alarm Low Limit (ART, P1, P2) 90, 50, 70 70, 40, 50 55, 20, 35
Alarm High Limit (PA) 35, 16, 20 60, 4, 26 60, 4, 26
Alarm Low Limit (PA) 10, 0, 0 24, -4, 12 24, -4, 12