Patient Monitor User Manual Alarms
- 46 -
In Menu > Maintenance > User Maintain > Alarm Setup, SpO
2
Sensor Off and ECG Lead
Off alarm level can be adjusted as High, Med. or Low. These alarm levels are defaulted to be
Low.
4.6 Network Disconnected Alarms
To configure the network disconnected alarms, select Menu > Maintenance > User Maintain >
Alarm Setup and choose Disconnect Alarm which can be set to On or Off. The alarm is Off by
default.
NOTE:
1 When the monitor is connected with the central monitoring system, you must set
Disconnect Alarm to On.
2 If Disconnect Alarm occurs during audio alarm paused or audio alarm off status, the
monitor will prompt a sounding alarm with information of Network Disconnect.
During the network disconnected status, activating audio alarm paused or audio
alarm off function can disable the audio alarm signal of Disconnect Alarm.
4.7 Testing Alarms
When you switch the monitor on, the monitor will prompt a “Di” tone that means the audio in
selftest is normal. Meantime, you must check that the alarm indicator lights are normal. This
indicates that the visible and audible alarm indicators are functioning correctly. For further testing
of individual measurement alarms, perform the measurement on yourself or use a simulator.
Adjust alarm limits and check that appropriate alarm behavior is observed.