EasyManuals Logo

EDAN elite V6 User Manual

EDAN elite V6
316 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #150 background imageLoading...
Page #150 background image
Patient Monitor User Manual Monitoring CO
2
- 135 -
15.2 CO
2
Safety Information
WARNING
1 Do not use the device in the environment with flammable anesthetic gas.
2 The device should be used by trained and qualified medical personnel authorized by
EDAN.
3 Nitrous oxide, elevated levels of oxygen, helium, xenon, halogenated hydrocarbons,
and barometric pressure can influence the CO
2
measurement.
4 The monitor will be damaged if any pipeline from the CO
2
module’s air tube /the air
inlet /the air outlet is plugged by water or other materials.
5 The accuracy of the CO
2
measurement will be affected by the following reasons: the
airway was highly obstructed; the leakage of air way connection or quick variation of
environment temperature.
6 Follow precautions for electrostatic discharge (ESD) and electromagnetic interference
(EMI) to and from other equipment.
7 Do not place the sensor cables or tubing in any manner that may cause entanglement
or strangulation.
8 When using mechanical ventilation, gas compensation should be well set.
Inappropriate setting may cause incorrect measurement result.
9 Respironics module is not equipped with automatic air pressure compensation, before
you start the CO
2
measurement for the first time, you must set the correct altitude.
Incorrect altitude settings can cause incorrect CO
2
readings. EDAN EtCO
2
module is
equipped with automatic air pressure compensation, and manual setting is not
required.
10 Leakage in the respiratory system or sampling system may result in a significant low
display of the EtCO
2
value. Always keep all components connected firmly and check
for leaks according to standard clinical procedures.
11 The EtCO
2
reading is not always closely related to the paCO
2
value, especially in
neonatal patients, and patients with pulmonary disease, with pulmonary embolism or
inappropriate ventilation.
12 Don’t measure CO
2
while nebulized medications are being delivered.
13 The CO
2
module temporally stops measuring during zeroing.
NOTE:
1 After the low battery alarm appears, please do not start the CO
2
measurement, or the
monitor may turn off for the low capacity of battery.
2 For disposal of hospital waste such as accumulated fluids, calibration gases,
sampled gases, where not otherwise specified, follow local regulations regarding
disposal of hospital waste.
3 If the measurement or sensor fails, stop measurement before the qualified service
personnel solves the problem.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EDAN elite V6 and is the answer not in the manual?

EDAN elite V6 Specifications

General IconGeneral
BrandEDAN
Modelelite V6
CategoryMedical Equipment
LanguageEnglish

Related product manuals