Patient Monitor User Manual Other Functions
- 213 -
situations such as frequent network disconnection may occur on the monitor.
5 The wireless driver is compatible with channels 1-11 only.
6 When signal intensity is level 2 or less, signal may be unstable and quality of the
signal transmission may be degraded.
7 When the monitor is connected to MFM-CMS via the wireless network, the user
should set the router to a secure encryption/authentication mode (Recommended
option: WPA2-PSK, with a high complexity, non-dictionary password).
28.3 Storing Data in the Storage Device
28.3.1 Data Stored in the Storage Device
Refer to Section A.9 Data Management for more information about single patient data volume.
When the single patient data reach the maximum, you can choose to
Keep Storing or Stop
Storing by selecting Menu > Common Function > Data Store > if one patient data full.
If you choose
Keep Storing, as soon as the single patient data is full, the earliest data will be
replaced by the latest one.
If you choose
Stop Storing, the monitor will stop data storing and the latest data cannot be stored
when the single patient data reach the maximum. For instance, if all the patient data (such as the
trend graph, trend table, NIBP measurements, arrhythmia event, alarm event and 12-lead analysis)
except waveforms reach the maximum, the monitor will stop storing, while only the waveforms
keep storing until they are full.
NOTE:
1 The storage time varies according to the patient's parameter data volume. When the
single patient data storage reaches 240 hours, the monitor will automatically create a
new folder for continuous data store.
2 When the storage space is insufficient, the monitor will stop storing data no matter
the user selects Keep Storing or Stop Storing.
28.3.2 Activating/ Deactivating Data Storing
To activate/ deactivate the data storing function, select Menu > Maintenance > User
Maintain > Other Setups, and set Data Store to On or Off.
The monitor will stop storing data in the storage device under the following circumstances:
No storage devices are selected.
There is no enough space in the storage device for storing data.
The removable device is read-only.
The data storing function is deactivated.
The monitor is switched off.
The power supply is off.