Patient Monitor User Manual Product Specifications
- 258 -
Resolution
0.1
°C (0.1 ºF)
☆Accuracy
1
± 0.3 °C
Refresh Time Every 1 s to 2 s
Temperature Calibration At an interval of 5 to 10 minutes
Measuring Mode Direct Mode
Transient Response Time ≤ 30 s
Note 1: The accuracy consists of two parts, as following:
Accuracy (not including sensor): ± 0.1 °C
Sensor accuracy: ≤ ± 0.2 °C
A.16 PR
Measuring range Accuracy Resolution
PR (SpO
2
) EDAN 25 bpm to 300 bpm
2 bpm
1 bpm
Nellcor 20 bpm to 300 bpm
3 bpm (20 bpm to 250
bpm)
1 bpm
PR
(NIBP)
EDAN 40 bpm to 240 bpm ±3 bpm or 3.5%, whichever
is greater
1 bpm
Omron Adult/ Pediatric mode: 40
bpm to 200 bpm
Neonatal mode: 40 bpm
to 240 bpm
2 bpm or 2% of the
readings
1 bpm
SunTech 30 bpm to 220 bpm ±3 bpm or ±2%, whichever
is greater
1 bpm
PR (IBP) EDAN 20 bpm to 300 bpm
30 bpm to 300 bpm:
±2 bpm
or
±2%, whichever is
greater;
20 bpm to 29 bpm:
undefined
1 bpm
A.17 IBP
Complies with IEC 60601-2-34: 2011.
Technique Direct invasive measurement
Channel 8