Patient Monitor User Manual Default Settings
- 292 -
Alarm High Limit (CVP/RAP/LAP/ICP) 10 4 4
Alarm Low Limit (CVP/RAP/LAP/ICP) 0 0 0
C.10 CO
2
Default Settings
CO
2
Settings ADU PED NEO
Alarm Switch On
Alarm Record Off
Alarm Level Medium
Work Mode Standby
Unit mmHg
Apnea Time 20 s
O
2
Compensate 16%
Anes Agent 0%
Alarm High Limit (EtCO
2
) 50 50 45
Alarm Low Limit (EtCO
2
) 25 25 30
Alarm High Limit (FiCO
2
) 4 4 4
Alarm High Limit (AwRR) 30 30 100
Alarm Low Limit (AwRR) 8 8 30
Sweep 6.25 mm/s
Amplitude Low
C.11 C.O. Default Settings
C.O. Settings ADU PED NEO
Alarm Switch On
Alarm Record Off
Alarm Level Medium
Alarm High Limit (TB) 40 40 40
Alarm Low Limit (TB) 30 30 30
Injective Temperature Source Auto
Temperature Unit
°C
Interval 30
Constant 0.542

Table of Contents