Patient Monitor User Manual Using Battery
- 220 -
4. The running time of the battery reflects the battery performance.
If the running time is obviously less than the specified time in the specification, please change the
battery or contact the service personnel.
29.5 Replacing the Battery
To install or replace the battery, please follow the procedure:
1.
To open the battery door, press the battery compartment latch and pull the battery door
according to indication beside the button.
2.
Remove the battery from the compartment.
3.
Insert a new battery into the battery compartment.
4.
Close the battery door.
NOTE:
The markers which respectively indicate compartment 1 and compartment 2 on the
battery door are corresponding to the symbols and on the main screen.
29.6 Recycling the Battery
When the battery no longer holds a charge, it should be replaced. Remove the old battery from
the monitor and recycle it properly.
WARNING
Do not disassemble batteries, put them into fire or cause them to short circuit. They may
ignite, explode or leak, causing personal injury.
29.7 Maintaining the Battery
To prolong the life of the batteries, there is current limitation for using batteries. Therefore, the
monitor which runs on battery power may not be powered on under following circumstances:
1.
Only one battery is installed.
2.
One of the two installed batteries is damaged, or large capacity difference between the two
installed batteries exists.
3.
Batteries in the monitor are almost empty.
If above-mentioned circumstances are detected, recharge the batteries or use another two batteries
with similar capacity.
CAUTION
1 Batteries should be conditioned regularly to maintain their useful life.
2 Remove the batteries from the monitor if they are not used for a longer period of time.
And recharge the batteries at a minimum of every 6 months when they are stored.
3 Discharge the battery completely once every month.