Patient Monitor User Manual Default Settings
- 294 -
Unit /
BIS Alarm High Limit 70
BIS Alarm Low Limit 20
C.14 RM Default Settings
RM Settings ADU PED NEO
Alarm Switch On
Alarm Level Medium
Alarm Record Off
Apnea Time 20 s
TV/MV TV
Respiration Mode Self-breath
Flow/Vol Flow
RR Alarm High Limit 30 30 60
RR Alarm Low Limit 8 8 30
PEEP Alarm High Limit 25 25 25
PEEP Alarm Low Limit 1 1 1
PIP Alarm High Limit 40 25 20
PIP Alarm Low Limit 1 1 1
MVe Alarm High Limit 8 4 0.8
MVe Alarm Low Limit 4 2.5 0.4
Loop Type P-V
Reference Loop On
Paw Ruler
Top Ruler: 40
Bottom Ruler: -40
Top Ruler: 40
Bottom Ruler:
-40
Top Ruler: 20
Bottom Ruler: -20
Vol Ruler
Top Ruler: 800
Bottom
Ruler:-800
Top Ruler: 200
Bottom
Ruler:-200
Top Ruler: 50
Bottom Ruler:-50
Flow Ruler
Top Ruler: 150
Bottom
Ruler:-150
Top Ruler: 100
Bottom
Ruler:-100
Top Ruler: 20
Bottom Ruler:-20