Patient Monitor User Manual Alarm Information
- 71 -
Message Cause Alarm Level Action Taken
Battery1 Current
Too High
Battery 1: discharge
over-current
Low
Stop using the
battery and notify
the manufacturer’s
service staff.
Battery2 Current
Too High
Battery 2: discharge
over-current
Low
Battery1 Charge
Voltage Too High
Battery 1: over-voltage
during charging
Low
Battery2 Charge
Voltage Too High
Battery 2: over-voltage
during charging
Low
Batter
Error in
iM20
Battery module
communication failure
Low
Printer Unavailable
The selected printer is not
available.
Low
Check whether the
network connection
is in good condition
and whether the
printer is
malfunctioning.
Insufficient storage
space
Less than 10 M space is left
in the storage device.
Low
Delete some data in
the storage device
or use another
removable device.
Read-only storage
device
The storage device is
read-only.
Low
Repair the storage
device or replace it
with a new one.
Audio Failed
Audio circuit connection is
abnormal, or loudspeaker
falls off.
High
Stop using the
monitor and notify
the manufacturer’s
service staff.
Network Disconnect
In distributed alarm system,
the monitor’s network is
disconnected.
Low
1) Check if the
network cable is
well connected.
2) Check if the
CMS is turned on.
3) Check if the IP of
bedside monitor and
CMS are on the
same network
segment.
5.3 Prompts
Message Cause
V-Fib/V-Tach Off
V-Fib/V-Tach alarm is set to Off.