Patient Monitor User Manual Alarm Information
- 53 -
Message Cause Alarm Level
FiSEV Low
FiSEV measuring value is below lower alarm limit. User-selectable
EtDES High
EtDES measuring value is above upper alarm limit. User-selectable
EtDES Low
EtDES measuring value is below lower alarm limit. User-selectable
FiDES High
FiDES measuring value is above upper alarm limit. User-selectable
FiDES Low
FiDES measuring value is below lower alarm limit. User-selectable
AG FiO
2
Low
FiO
2
measure value is below 18%. High
AG APNEA
In the set apnea alarm delay time interval, no breath
can be detected using AG module.
High
TB High
TB measuring value is above upper alarm. User-selectable
TB Low
TB measuring value is below lower alarm. User-selectable
BIS High
BIS measuring value is above upper alarm. User-selectable
BIS Low
BIS measuring value is below lower alarm. User-selectable
RM Apnea
In a specific time interval, no respiration can be
detected by RM module.
High
AwRR (RM)
High
AwRR (RM) measuring value is above upper alarm
limit.
User-selectable
AwRR (RM)
Low
AwRR (RM) measuring value is below lower alarm
limit.
User-selectable
PEEP High
PEEP measuring value is above upper alarm limit. User-selectable
PEEP Low
PEEP measuring value is below lower alarm limit. User-selectable
PIP High
PIP measuring value is above upper alarm limit. User-selectable
PIP Low
PIP measuring value is below lower alarm limit. User-selectable
MVe High
MVe measuring value is above upper alarm limit. User-selectable
MVe Low
MVe measuring value is below lower alarm limit. User-selectable
CI High
CI measuring value is above upper alarm limit. User-selectable
CI Low
CI measuring value is below lower alarm limit. User-selectable
TOF High
TOFrat/TOFcnt measuring value is above upper alarm
limit.
User-selectable
TOF Low
TOFrat/TOFcnt measuring value is below lower alarm
limit.
User-selectable
5.2 Technical Alarm Information
NOTE:
The ECG alarm information listed in the below table describes the lead names in
America. For the corresponding lead names in Europe, please refer to the section
Installing Electrodes.