Patient Monitor User Manual Product Specifications
- 277 -
Dynamic
Compliance (Cdyn)
Adult:
10.0 mL/cmH
2
O to 100.0
mL/cmH
2
O
Pediatric:
5.0 mL/cmH
2
O to 50
mL/cmH
2
O
Neonatal: 1.0 mL/cmH
2
O to
15 mL/cmH
2
O
0.1 mL/cmH
2
O
EtCO
2
(CO
2
sensor
is required)
5.0 mmHg to 150.0 mmHg
(0.7 kPa to 20.0 kPa/ 0.7% to
19.7%)
1 mmHg
FiCO
2
(CO
2
sensor
is required)
3.0 mmHg to 50.0 mmHg
(0.4 kPa to 6.6 kPa/ 0.4% to
6.6%)
1 mmHg
☆Alarm Type
AwRR, PIP, PEEP, MVe
☆ Apnea alarm
delay
10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s; default value is 20 s.
A.23 ICG
Technique Thoracic electrical bioimpedance
☆ Measuring
Range
SV: 0 ml/beat~250 ml/beat
HR: 40 bpm
~250 bpm
C.O.: 0 L/min
~30 L/min
☆Accuracy
SV: Undefined
HR: ± 2 bpm
C.O.: Undefined
A.24 LiDCO
Measure
Parameters
CO, CI, SV, SI, SVR, SVRI, DO
2
, DO
2
I, SaO
2
, CVP, HR, MAP, DIA,
SYS, SVV, PPV, SPV and HRV.
A.25 Interfaces
A.25.1 Analog Output
Bandwidth (-3 dB; reference
frequency: 10 Hz)
Diagnosis/Monitor: 0.5 Hz to 40 Hz
Surgery: 1 Hz to 20 Hz