Patient Monitor User Manual EMC Information
- 281 -
B EMC Information
- Guidance and Manufacture’s Declaration
B.1 Electromagnetic Emissions
Guidance and manufacture’s declaration – electromagnetic emission
elite V5 V6 V8 is intended for use in the electromagnetic environment specified below. The
customer or the user of elite V5 V6 V8 should assure that it is used in such an environment.
Emission test Compliance Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
elite V5 V6 V8 uses RF energy only for its
internal function. Therefore, its RF emissions are
very low and are not likely to cause any
interference in nearby electronic equipment.
RF emission
CISPR 11
Class A
elite V5 V6 V8 is suitable for use in all
establishments, other than domestic
establishments and those directly connected to the
public low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic emissions
IEC/EN 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC/EN 61000-3-3
Complies
NOTE:
The EMISSIONS characteristics of elite V5 V6 V8 make it suitable for use in industrial
areas and hospitals (CISPR 11 class A). If it is used in a residential environment (for
which CISPR 11 class B is normally required) iM50/iM60/iM70/iM80/M50/M80 might not
offer adequate protection to radio-frequency communication services. The user might
need to take mitigation measures, such as relocating or re-orienting the equipment.
B.2 Electromagnetic Immunity
Guidance and manufacture’s declaration – electromagnetic immunity
elite V5 V6 V8 is intended for use in the electromagnetic environment specified below. The
customer or the user of elite V5 V6 V8 should assure that it is used in such an environment.
Immunity test
IEC/EN 60601 test
level
Compliance level
Electromagnetic
environment - guidance

Table of Contents