Patient Monitor User Manual EMC Information
- 282 -
Electrostatic
discharge (ESD)
IEC/EN 61000-4-2
± 8 kV contact
± 15 kV air
8 kV contact
± 15 kV air
Floors should be wood,
concrete or ceramic tile.
If floor are covered with
synthetic material, the
relative humidity should
be at least 30%.
Electrical fast
transient/burst
IEC/EN 61000-4-4
± 2 kV for power supply
lines
2 kV for power
supply lines
Mains power quality
should be that of a
typical commercial or
hospital environment.
Surge
IEC/EN 61000-4-5
± 1 kV for line to line
± 2 kV for line to
ground
1 kV for line to
line
± 2 kV for line to
ground
Mains power quality
should be that of a
typical commercial or
hospital environment.
Power frequency
(50 Hz /60 Hz)
magnetic field
IEC/EN 61000-4-8
30 A/m 30 A/m Power frequency
magnetic fields should
be at levels characteristic
of a typical location in a
typical commercial or
hospital environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC/EN
61000-4-11
0 % U
T;
0.5 cycle
At 0°, 45°, 90°, 135°,
180°, 225°, 270° and
315°
0 % U
T
; 1 cycle
and
70 % U
T
; 25/30 cycles )
Single phase: at 0°
0 % U
T
; 250/300 cycle
0 % U
T;
0.5 cycle
At 0°, 45°, 90°,
135°, 180°, 225°,
270° and 315°
0 % U
T
; 1 cycle
and
70 % U
T
; 25/30
cycles )
Single phase: at 0°
0 % U
T
; 250/300
cycle
Mains power quality
should be that of a
typical commercial or
hospital environment. If
the user of elite V5 V6
V8 requires continued
operation during power
mains interruptions, it is
recommended that elite
V5 V6 V8 be powered
from an uninterruptible
power supply or a
battery.
NOTE U
T
is the a.c. mains voltage prior to application of the test level.