Patient Monitor User Manual Alarm Information
- 70 -
Message Cause Alarm Level Action Taken
NMT Temp. Value
Out Of Range
External device’s temp
sensor signal over flow and
exceed rated measurement
range.
Low
Make sure temp
sensor is well
connected and
temperature is
within measurement
range,
PTC error
In PTC mode, the patient
response in the first
stimulation.
Low
Check the patient
response and drug
dose is proper.
Other external devices
XX Comm. Failed
(XX represents one of
the external devices
connected.)
External module failure or
communication failure.
High Unplug the V-Link
module, insert the
serial port cable, and
then reinsert V-Link
module. If the
problem still exists,
please notify
biomedical engineer
or manufacturer’s
service staff.
V-Link Abnormity
Communication failure
between monitor and
V-Link.
Medium Please contact the
manufacturer and
replace V-Link
module.
Others
Battery Low
Battery Low
High
Change the batteries
or charge the
batteries.
Battery1 Error
Malfunction in Battery 1
Low
Replace the battery
and restart the
monitor. If the
problem persists,
notify the
manufacturer’s
service staff.
Battery2 Error
Malfunction in Battery 2
Low
Recorder Out Of
Paper
Recorder Out Of Paper Low
Please install the
paper
Recorder Probe
Overheated
The probe of recorder is
overheated.
Low
Stop recording and
retry after the probe
cools.