Patient Monitor User Manual Alarm Information
- 69 -
Message Cause Alarm Level Action Taken
NMT Battery
Depleted
External device battery is
depleted.
Medium
Replace or charge
the battery.
Skin resistance too
high
Because electrode position is
wrong, electrode is not well
fixed on skin or patient skin
resistance problem,
resistance is too high, and
then measurement can’t be
taken.
Medium
Check the electrode
and cable to confirm
connection to
patient, the monitor
and external device
cable are well
connected.
NMT Calibration
Failed
Electrode position is wrong,
frequency and sensitivity is
not proper for patient type.
Medium
Check the electrode
and patient setup to
make sure electrode
is well connected
and patient setup is
correct.
NMT Internal Err.
External device’s internal
hardware or system error
occurs.
Medium
Check external
device.
NMT Comm. Fail
V-Link module
communication failure.
High
Stop measuring
function of NMT
module, and notify
biomedical engineer
or manufacturer’s
service staff.
NMT Surface/Needle
Cable Not Mounted
Surface cable and needle
cable are both not mounted.
Low
Connect surface
cable and needle
cable
NMT Acc. Sensor
Cable Not Mounted
External device’s acc cable
is not mounted.
Low
Connect sensor
cable.
NMT Temp. Cable
Not Mounted
Temp Cable is not mounted. Low
Connect TEMP
cable.
NMT Acc. Sensor
Signal Out Of Range
External device’s acc sensor
signal is too high to exceed
rated measurement range.
Low
Make sure acc
sensor electrode
connection is well,
and reduce patient
finger motion.
NMT Battery Low
External device battery is
low.
Medium
Replace or charge
the battery.