Patient Monitor User Manual Alarm Information
- 68 -
Message Cause Alarm Level Action Taken
ICG L1 Lead Off
The No.1 lead on the left is
off.
Low
Make sure the No.1
lead on the left is
properly connected.
ICG R1 Lead Off
The No.1 lead on the right is
off.
Low
Make sure the No.1
lead on the right is
properly connected.
ICG L2 or L3 Lead
Off
The No.2 or No.3 lead on
the left is off.
Low
Make sure the No.2
or No.3 lead on the
left is properly
connected.
ICG R2 or R3 Lead
Off
The No.2 or No.3 lead on
the right is off.
Low
Make sure the No.2
or No.3 lead on the
right is properly
connected.
ICG L4 Lead Off
The No.4 lead on the left is
off.
Low
Make sure the No.4
lead on the left is
properly connected.
ICG R4 Lead Off
The No.4 lead on the right is
off.
Low
Make sure the No.4
lead on the right is
properly connected.
ICG Comm Fail
The communication
between the ICG module
and the monitor fails during
measuring.
High
Unplug the module
and plug it again. If
the problem still
exists, contact your
service personnel.
NMT (TOF Watch)
NMT Stim. Signal
Missing
During stimulation, the
electrode is moved, dropped
or broken. The monitor or
external device is
disconnected with the
stimulation cable.
Medium
Check the electrode
and cable to confirm
connection to
patient, the monitor
and external device
cable are well
connected.
NMT Acc. Signal
Missing
During stimulation, acc
sensor is moved, dropped or
broken. Or in calibration
mode, acc sensor connection
is not well or signal is too
weak.
Medium