Patient Monitor User Manual Alarm Information
- 67 -
Message Cause Alarm Level Action Taken
CO
2
(RM) Check
Adapter
The cannula is off or
disconnected.
Low
Check whether the
adapter is properly
connected or
replace the adapter.
CO
2
(RM) Sensor
Faulty
CO
2
module failure High
Stop using
measuring function
of CO
2
module,
notify biomedical
engineer.
CO
2
(RM) Sensor
Over Temp
CO
2
measure value exceeds
the measure range of the
monitor.
High
CO
2
(RM) Zero
Required
Zero calibration failure Low
Disconnect the
sampling cannula or
adapter from the
airway; initiate the
zeroing before
making sure that no
expired air is inside
the sampling
cannula and adapter.
CO
2
(RM) Out Of
Range
The CO
2
concentration
exceeds the accuracy range
of RM module.
High
Reduce CO
2
concentration.
CO
2
(RM) Sensor
Off
The sensor may be
disconnected from the
patient or the monitor.
Low
Check the sensor
connection.
ICG
ICG Sensor Off
1) The ICG sensor is
disconnected from the
module
2) Bad connection
Low
Reconnect the ICG
sensor.
ICG Need Input
Param
The required patient data,
such as height and weight,
has not been input into the
monitor.
Low Input patient data.
ICG Input Param
Error
The input patient data is
invalid.
Low
Input valid patient
data.