Patient Monitor User Manual Product Specifications
- 254 -
Neonatal (147 ± 3) mmHg
Pre-inflation Pressure
Adult Mode Default: 160 mmHg
Range: 80/100/120/140/150/160/180/200/220/240 mmHg
Pediatric Mode Default: 140 mmHg
Range: 80/100/120/140/150/160/180/200 mmHg
Neonatal Mode Default: 100 mmHg
Range: 60/70/80/100/120 mmHg
Omron Module
Method Oscillometric
Mode Manual, Auto, Continuous
Measuring Interval in AUTO Mode 1/2/3/4/5/10/15/30/60/90 min, 2/4/8 h
Continuous 5 min, interval is 5 s
☆Measuring Type
SYS, DIA, MAP
☆Measuring Range
☆Adult/ Pediatric Mode
SYS: 60 mmHg to 250 mmHg
DIA: 40 mmHg to 200 mmHg
MAP: 45 mmHg to 235 mmHg
☆Neonatal Mode
SYS: 40 mmHg to 120 mmHg
DIA: 20 mmHg to 90 mmHg
MAP: 30 mmHg to 100 mmHg
Alarm Type SYS, DIA, MAP
Cuff pressure Measuring Range 0 mmHg to 300 mmHg
Pressure Resolution 1 mmHg
Measuring Accuracy
☆Maximum Mean Error
±5 mmHg
☆ Maximum Standard
Deviation
8 mmHg
Adult/Pediatric Neonate
Normal
Condition
Single Fault
Condition
Normal
Condition
Single fault
Condition