EasyManua.ls Logo

EDAN elite V6 - Page 269

EDAN elite V6
316 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Patient Monitor User Manual Product Specifications
- 254 -
Neonatal (147 ± 3) mmHg
Pre-inflation Pressure
Adult Mode Default: 160 mmHg
Range: 80/100/120/140/150/160/180/200/220/240 mmHg
Pediatric Mode Default: 140 mmHg
Range: 80/100/120/140/150/160/180/200 mmHg
Neonatal Mode Default: 100 mmHg
Range: 60/70/80/100/120 mmHg
Omron Module
Method Oscillometric
Mode Manual, Auto, Continuous
Measuring Interval in AUTO Mode 1/2/3/4/5/10/15/30/60/90 min, 2/4/8 h
Continuous 5 min, interval is 5 s
Measuring Type
SYS, DIA, MAP
Measuring Range
Adult/ Pediatric Mode
SYS: 60 mmHg to 250 mmHg
DIA: 40 mmHg to 200 mmHg
MAP: 45 mmHg to 235 mmHg
Neonatal Mode
SYS: 40 mmHg to 120 mmHg
DIA: 20 mmHg to 90 mmHg
MAP: 30 mmHg to 100 mmHg
Alarm Type SYS, DIA, MAP
Cuff pressure Measuring Range 0 mmHg to 300 mmHg
Pressure Resolution 1 mmHg
Measuring Accuracy
Maximum Mean Error
±5 mmHg
Maximum Standard
Deviation
8 mmHg
Adult/Pediatric Neonate
Normal
Condition
Single Fault
Condition
Normal
Condition
Single fault
Condition

Table of Contents