Patient Monitor User Manual Product Specifications
- 255 -
Maximum Cuff Pressure 300 mmHg 330 mmHg 150 mmHg 165 mmHg
Maximum Measuring
Period
Less than
160 s
Less than 180 s Less than 80 s Less than 90
s
Pre-inflation Pressure
Adult/ Pediatric Mode Default: 180 mmHg
Range: 120/140/150/160/180/200/220/240/260/280 mmHg
Neonatal Mode Default: 120 mmHg
Range: 80/100/120/140 mmHg
Dual Independent Channel Overpressure Protection
Adult/Pediatric < 300 mmHg
Neonatal < 150 mmHg
SunTech Module
Method Oscillometric
Mode Manual, Auto, Continuous
Measuring Interval in AUTO Mode 1/2/3/4/5/10/15/30/60/90/120/240/480 (unit: minute)
☆Measuring Type
SYS, DIA, MAP
☆Measuring Range
☆Adult Mode
SYS: 40 mmHg ~ 260 mmHg
DIA: 20 mmHg ~ 200 mmHg
MAP: 26 mmHg ~ 220 mmHg
☆Pediatric Mode
SYS: 40 mmHg ~ 230 mmHg
DIA: 20 mmHg ~ 160 mmHg
MAP: 26 mmHg ~ 183 mmHg
☆Neonatal Mode
SYS: 40 mmHg ~ 130 mmHg
DIA: 20 mmHg ~ 100 mmHg
MAP: 26 mmHg ~ 110 mmHg
☆Alarm Type
SYS, DIA, MAP
Pressure Resolution 1 mmHg
☆Maximum mean error
±5 mmHg
☆Maximum standard deviation
8 mmHg
Maximum measuring period