EasyManua.ls Logo

EDAN elite V6 - Page 270

EDAN elite V6
316 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Patient Monitor User Manual Product Specifications
- 255 -
Maximum Cuff Pressure 300 mmHg 330 mmHg 150 mmHg 165 mmHg
Maximum Measuring
Period
Less than
160 s
Less than 180 s Less than 80 s Less than 90
s
Pre-inflation Pressure
Adult/ Pediatric Mode Default: 180 mmHg
Range: 120/140/150/160/180/200/220/240/260/280 mmHg
Neonatal Mode Default: 120 mmHg
Range: 80/100/120/140 mmHg
Dual Independent Channel Overpressure Protection
Adult/Pediatric < 300 mmHg
Neonatal < 150 mmHg
SunTech Module
Method Oscillometric
Mode Manual, Auto, Continuous
Measuring Interval in AUTO Mode 1/2/3/4/5/10/15/30/60/90/120/240/480 (unit: minute)
Measuring Type
SYS, DIA, MAP
Measuring Range
Adult Mode
SYS: 40 mmHg ~ 260 mmHg
DIA: 20 mmHg ~ 200 mmHg
MAP: 26 mmHg ~ 220 mmHg
Pediatric Mode
SYS: 40 mmHg ~ 230 mmHg
DIA: 20 mmHg ~ 160 mmHg
MAP: 26 mmHg ~ 183 mmHg
Neonatal Mode
SYS: 40 mmHg ~ 130 mmHg
DIA: 20 mmHg ~ 100 mmHg
MAP: 26 mmHg ~ 110 mmHg
Alarm Type
SYS, DIA, MAP
Pressure Resolution 1 mmHg
Maximum mean error
±5 mmHg
Maximum standard deviation
8 mmHg
Maximum measuring period

Table of Contents

Related product manuals