Patient Monitor User Manual Monitoring C.O.
- 146 -
Window for C.O. Edit
Contents displayed in the window:
Six curves of the six measurements and C.O. value
Average value of C.O.
Average value of CI
WARNING
1 Make sure that the computational constant for the measurement is appropriate to the
catheter used.
2 Before a C.O. measurement is initiated, check the accuracy of patient setup. The
calculation of C.O. is related to the patient height, weight, and catheter computation
coefficient; therefore, incorrect input will lead to error in calculation.
NOTE:
1 The blood temperature alarm will not function during C.O. measurement. It will
resume automatically when the measurement is over.
2 It is strongly recommended that the user must push the injector within four seconds
after pressing the Start button.
3 It is strongly recommended that you wait at least 1 minute (or longer depending on
the patient’s clinical condition) before starting the next measurement.