Patient Monitor User Manual Product Specifications
- 251 -
Ventricular
Tachycardia
1 mV 206 bpm
Gain 0.5: 10 s
Gain 1.0: 10 s
Gain 2.0: 10 s
Ventricular
Tachycardia
2 mV 195 bpm
Gain 0.5: 10 s
Gain 1.0: 10 s
Gain 2.0: 10 s
Response Time of
Heart Rate Meter to
Change in HR
HR range: 80 bpm to 120 bpm
Range : Within 11 s
HR range: 80 bpm to 40 bpm
Range : Within 11 s
☆ Tall T-wave
Rejection
Complied with IEC 60601-2-27: 2011, Sect. 201.12.1.101.17
minimum recommended 1.2 mV T-Wave amplitude
Accuracy of Heart
Rate Meter and
Response to Irregular
Rhythm
Complied with IEC 60601-2-27: 2011, Sect. 201.7.9.2.9.101 b) 4), the
HR value after 20 seconds of stabilization is displayed as follows:
Ventricular bigeminy: 80 bpm ±1 bpm
Slow alternating ventricular bigeminy: 60 bpm ±1 bpm
Rapid alternating ventricular bigeminy: 120 bpm ±1 bpm
Bidirectional systoles: 91 bpm ±1 bpm
Time to Alarm for
Heart Rate alarm
conditions
Asystole alarm: ≤ 10 s
HR low alarm: ≤ 10 s
HR high alarm: ≤ 10 s
Arrhythmia analyses
Asystole V-Fib/V-Tach Couplet
Vent Rhythm PVC Bigeminy PVC Trigeminy
Tachy R on T PVC
Irr Rhythm Brady Missed Beat
Pacer not Pacing Vent Brady Pacer not Capture
VEB Run PVCs Acc. Vent Rhythm
IPVC Non-Sustain VT Multiform PVCs
Pauses/min High Pause Afib
PAC Bigeminy PVCs High Low Voltage(Limb)
ExtremeBrady PAC Trigeminy Wide QRS Tachy
Sustain VT ExtremeTachy V-Tach