Patient Monitor User Manual Product Specifications
- 262 -
Measure Parameters EtCO
2
, FiCO
2
, AwRR
Unit mmHg, %, Kpa
Measuring Range
EtCO
2
0 mmHg to 150 mmHg
FiCO
2
3 mmHg to 50 mmHg
AwRR
0 rpm to 150 rpm (Mainstream)
2 rpm to 150 rpm (Sidestream)
Resolution EtCO
2
1 mmHg
FiCO
2
1 mmHg
AwRR 1 rpm
EtCO
2
Accuracy
± 2 mmHg, 0 mmHg to 40 mmHg
± 5% of reading, 41 mmHg to 70 mmHg
± 8% of reading, 71 mmHg to 100 mmHg
± 10% of reading, 101 mmHg to 150 mmHg
± 12% of reading, RR is over 80 rpm (sidestream)
There will be no degradation in performance due to
Respiration Rate. (mainstream)
AwRR Accuracy
± 1 rpm
Operation Mode Measure, standby
Sample Gas Flowrate
(sidestream)
(50 ± 10) ml/min
O
2
Compensation
Range 0% to 100%
Resolution 1%
Default 16%
Barometric Pressure
Compensation
User setup
Anesthetic Gas Compensation
Range 0% to 20%
Resolution 0.1%
Default 0.0%