Patient Monitor User Manual Product Specifications
- 275 -
Subparameters
Parameters Range Resolution
Peak Inspiratory
Pressure (PIP)
1 cmH
2
O to 120.0 cmH
2
O
1 cmH
2
O
Plateau Pressure
(Pplat)
Adult/Pediatric: 1.0 cmH
2
O
to 99 cmH
2
O
1 cmH
2
O
Positive
End-Expiratory
Pressure (PEEP)
1.0 cmH
2
O to 50.0 cmH
2
O
1 cmH
2
O
Mean Airway
Pressure (Pmean)
0.0 cmH
2
O to 100.0 cmH
2
O
1 cmH
2
O
Peak Inspiratory
Flow (PIF)
Adult: 2.0 L/min to 180.0
L/min
Pediatric: 0.75 L/min to
100.0 L/Min
Neonatal: 0.25 L/min to
30.00 L/Min
Adult/Pediatric/ Neonatal: 1 L/min
Peak Expiratory
Flow (PEF)
Adult: 2.0 L/min to 180.0
L/min
Pediatric: 0.75 L/min to
100.0 L/min
Neonatal: 0.25 L/min to
30.00 L/min
Adult/Pediatric/ Neonatal: 1 L/min
Inspired Minute
Volume (MVi)
Adult: 1 L/min to 30.00
L/min
Pediatric: 0.3 L/min to 20
L/min
Neonatal: 0.1 L/min to 3
L/min
Adult/Pediatric/ Neonatal: 0.1 L/min
Expired Minute
Volume (MVe)
Adult: 1 L/min to 30.00
L/min
Pediatric: 0.3 L/min to 20
L/min
Neonatal: 0.1 L/min to 3
L/min
Adult/Pediatric/ Neonatal: 0.1 L/min