Patient Monitor User Manual Basic Operation
- 21 -
1 Power cable safety
latch
Used to prevent the power cable from loosing
or falling. Place the latch on the power cable
and press it down firmly to ensure that it
secures the power cable.
2 AC power input
3 Equipotential
grounding terminal
If the monitor is used together with other
devices, connect this terminal to eliminate
potential ground differences between devices.
4 RS232 interface Connect it to communicate with other devices.
5 USB interfaces They support USB1.0/2.0 output.
6 Nurse call port/ analog
output/ defibrillator
synchronization
Nurse call port: it connects the monitor to the
hospital’s nurse call system. Alarms
indications are alerted through the nurse call
system if configured to do so.
Analog output: the monitor outputs the
waveform through the port.
Defibrillator synchronization: the monitor
outputs the defibrillator synchronization signal
through the port.
7 VGA output It enables the VGA video output.
8 PAM connector It connects the Parameter Amplifier
Mainframe to the monitor.
9 Network interface It connects the monitor to the central
monitoring system (MFM-CMS) or gateway
via standard network cable, which enables
MFM-CMS or gateway to achieve
bidirectional communication with the monitor.
10 Extended video
interface
It connects a secondary display, which extends
the display capability of your monitor.
11 Anti-thief lock